FAQ's

Yes, this system will work with all three formats: film, phosphor plates and sensors.

Yes, the system has various arm length options to suit your surgery setup.

Yes, the system is has simple controls and pre-set exposure times depending on the location of the x-ray in the patients dentition

Yes, the x-ray will be tested by our engineer after installation.


Maintenance & Aftercare

In order to guarantee safety of the X-ray system, a maintenance schedule must be put in place. The owner is responsible for setting up and adhering to a schedule for maintenance which must be carried out by qualified technicians who are able to certify their work with a “Compliance Declaration”.

Run a check on the system and its operation when it is installed and every twelve months thereafter.

Once a year, lubricate the pins and bushes of the wall plate and positioning arm, as specified (refer to the Installation & Maintenance Manual)

Do not lose the adjustment key that comes with the system, since, in time, it might become necessary to make readjustments.

If parts become hard to move or squeak, call our service team on 0800 028 4749.

Use a damp soft cloth and soapy water to clean the outer surfaces.

The spacer cone may be cleaned with cotton wool and surgical alcohol.


Troubleshooting

Detailed below is a list of error messages which may appear on your unit. Click on a message to see what this means and what action should be taken.

If you do not find the problem below or the action you take does not solve the problem, please contact our support team on 08700 10 20 46. In order to help you, have the serial number of your unit and a description of the error.

Possible Cause

RX1 tubehead is not connected or is out of order.

Action

Call our service team on 0800 028 4749.

Possible Cause

RX2 tubehead is not connected or is out of order.

Action

Call our service team on 0800 028 4749.

Possible Cause

Corrupted eeprom data.

Action

Call our service team on 0800 028 4749.

Possible Cause

Eeprom data not saved properly.

Action

Call our service team on 0800 028 4749.

Possible Cause

Line voltage value not within set limits.

Action

Call our service team on 0800 028 4749.

Possible Cause

Line voltage value not within the ± 15% nominal value.

Action

Call our service team on 0800 028 4749.

Possible Cause

The "x-ray" key always seems to be pressed.

Action

Ensure it is not jammed.

Possible Cause

Anomaly in the control panel.

Action

Call our service team on 0800 028 4749.

Possible Cause

The exposure has been prematurely interrupted.

Action

Keep the "x-ray" key pressed until the end of the exposure.

Possible Cause

Anomaly in the Triac/Relay.

Action

Call our service team on 0800 028 4749.

Possible Cause

Anomaly in the electronic circuit.

Action

Call our service team on 0800 028 4749.

Possible Cause

Anomaly in the control circuit.

Action

Call our service team on 0800 028 4749.

Possible Cause

Incorrect dip-switch configuration.

Action

Call our service team on 0800 028 4749.

Possible Cause

The "Control Button" does not correspond to the selected tubehead.

Action

Select the "Control Button" that corresponds to the selected tubehead.

Possible Cause

The tubehead does not work properly.

Action

Call our service team on 0800 028 4749.

Possible Cause

The tubehead is not in the correct mode.

Action

Call our service team on 0800 028 4749.

Possible Cause

The tubehead has not completed the exposure.

Action

Repeat the exposure or call our service team on 0800 028 4749.

Possible Cause

Problem in the frequency or regulation.

Action

Call our service team on 0800 028 4749.

Possible Cause

The tubehead is not calibrated.

Action

Calibrate the tubehead or call our service team on 0800 028 4749.

Possible Cause

Eeprom data not saved properly.

Action

Call our service team on 0800 028 4749.

Possible Cause

Corrupted eeprom data.

Action

Call our service team on 0800 028 4749.

Possible Cause

Overvoltage error.

Action

Call our service team on 0800 028 4749.

Possible Cause

Anode voltage out of tolerance.

Action

Call our service team on 0800 028 4749.

Possible Cause

Anode current out of tolerance.

Action

Call our service team on 0800 028 4749.

Possible Cause

Control connector.

Action

Call our service team on 0800 028 4749.

Possible Cause

Problem in the reference voltage.

Action

Call our service team on 0800 028 4749.

Possible Cause

Major error.

Action

All x-ray system functions are disabled. Call our service team on 0800 028 4749