If used with care, imaging plates can be used indefinitely until physically damaged. However, the imaging plates will show wear over time due to continuous use. Inspect the imaging plates regularly and replace them if they are scratched and/or show signs of excessive wear.
To prevent damage to the imaging plates and the possibility of image artifacts, avoid contact between the imaging plates and the following materials/solutions/solvents: Isopropyl alcohol, hydrogen peroxide and other peroxides, citrus-based cleaners, hand lotions and water-less hand sanitizers, as well as surfactants and lubricants.
Handling and Storage of Imaging Plates
Cleaning the Imaging Plate
Clean the imaging plate to remove dirt, dust particles, finger prints and so forth that could affect image quality.
Cleaning Materials
Cleaning Method
To clean the imaging plates, follow these steps:
Do not soak the hygienic sheath in any cleaning or disinfecting solutions. Do not autoclave; autoclaved hygienic sheaths must be discarded.
Frequency of Disinfection
The hygienic sheath is the primary means of infection control and protection from contaminated imaging plates.
Disinfect the imaging plate after every use (after the image plate is scanned), before it is inserted in the hygenic sheath for the next use.
Disinfection Solutions
To disinfect, use either a self-prepared solution, or a commercially prepared product that adheres to the following specifications and restrictions:
Disinfection Method
If a commercially prepared equivalent solution of diluted bleach is used, follow the manufacturer’s specifications and warnings.
To disinfect the hygienic sheaths and imaging plates, follow these steps:
The imaging plate contains Barium and should be considered hazardous or special waste in specific conditions at the end of its useful service life. For disposal or recycling information, contact your local authorities.
Cleaning the Scanner Body : Use disinfection wipes for medical equipment to clean the scanner body thoroughly from all sides, including the tray. It is the user’s responsibility to disinfect the imaging plates tray daily to prevent cross contamination
Cleaning the Detachable Insertion Panels : The CS 7600 makes use of three insertion panel sizes according to the inserted plates size (0, 0–3, and 4). The maintenance procedure is identical for both insertion panels. It is the user’s responsibility to disinfect the detachable insertion panel(s) daily to prevent cross contamination. Do not autoclave; autoclaved detachable insertion panel(s) must be discarded.
To clean and disinfect the insertion panels, follow these steps:
Before you perform this procedure, make sure you have the rollers cleaning plate.
The scanner rollers should be cleaned in the following cases:
To clean the scanner rollers, follow these steps:
Cleaning the rollers requires using the size 4 insertion slot
Tip: The cleaning procedure is to be repeated 3 times to ensure the rollers are thoroughly cleaned.
Cleaning the Scan & Go : Use disinfection wipes for medical equipment to clean the Scan & Go device thoroughly from all sides.
Replacing the Battery (Scan & Go Generation 1) : For Scan & Go Generation 1 models released before June 2012. Scan & Go devices after June 2012 do not have a battery.
Replace only with a Carestream genuine BL-5C battery
To replace the Scan & Go battery, follow these steps:
Detailed below is a list of malfunctions which may appear on your scanner’s LCD and in the image acquisition interface. Click on a message to see the possible cause of the malfunction and what action should be taken.
If you do not find the problem below or the action you take does not solve the problem, please contact our support team on 08700 10 20 46. In order to help you, have the model number of your unit and details of the malfunction.
Image or Image on scanner display
And/or
Image in the image acquisition interface
Possible Cause
Limited or no connection to scanner (physical connection is intact, but the IP address is not correct).
Action
ImageImage appears.
Possible Cause
No connection to scanner after launching image acquisition interface.
Action
Possible Cause
Scanner memory full.
Action
Clear scanner memory.
Clearing the Scanner Memory
The scanner’s storage capacity is 25 images. Images are stored in the scanner’s memory in the following cases:
When the scanner’s memory nears its limit, the scan number icon changes to red.
To clear the scanner’s memory, follow these steps:
Possible Cause
Imaging plate loading failure.
Action
Possible Cause
Imaging plate inserted incorrectly.
Action
Important: Do not insert any tools into the scanner. If any tool is inserted into the scanner the warranties are null and void.
To retrieve a fallen imaging plate, follow these steps:
Important: If the plate is not in the location shown in the above picture, do not continue to look for the plate, call our service team on 0800 028 4749.
Possible Cause
Either imaging plate inserted incorrectly or the imaging plate hygienic sheath is inserted in the wrong direction.
Action
Possible Cause
The scanned image was routed to the scanner’s memory.
Action
Use the image Acquisition interface to check the selected scanner’s unclaimed images and retrieve the required images.
Possible Cause
Incompatible insertion panel size.
Action
Check if the insertion panel size is compatible to the imaging plate size.
Possible Cause
Scanner rollers have accumulated grime.
Action
Clean the scanner rollers.
Note: Before you perform this procedure, make sure you have the rollers cleaning plate.
The scanner rollers should be cleaned in the following cases:
To clean the scanner rollers, follow these steps:
Note: Cleaning the rollers requires using the size 4 insertion slot.
Symbol appears in the image acquisition interface display
Possible Cause
No USB connection to Scan & Go device.
Action