The HYDRIM C61W/WD G4 dental washer disinfector from SciCan is a high-performance washer-disinfector designed to streamline and optimise instrument reprocessing. With a capacity of six full-size cassettes and the ability to clean and dry 120 instruments in just 40 minutes, this efficient system saves time and ensures consistent, validated cleaning cycles.
Fast Track allows dental practices to acquire dental equipment quickly, with delivery possible within one week.
Say goodbye to long waiting times and say hello to a hassle-free experience.
Fast Track allows dental practices to acquire dental equipment quickly, with delivery possible within one week.
Say goodbye to long waiting times and say hello to a hassle-free experience.
Fast Track allows dental practices to acquire dental equipment quickly, with delivery possible within one week.
Say goodbye to long waiting times and say hello to a hassle-free experience.

If you're about to invest in dental equipment, you could be able to combine your costs and benefit from Advance.
Eligible customers will receive a rebate based on their merchandise and clear aligner spend, which can help cover the cost of your equipment.
If you're about to invest in dental equipment, you could be able to combine your costs and benefit from Advance.
Eligible customers will receive a rebate based on their merchandise and clear aligner spend, which can help cover the cost of your equipment.

If you're about to invest in dental equipment, you could be able to combine your costs and benefit from Advance.
Eligible customers will receive a rebate based on their merchandise and clear aligner spend, which can help cover the cost of your equipment.

Detailed below is a list of errors which may occur. Click on an error to see the possible cause and what action should be taken.
If you do not find the error below or the action you take does not solve the problem, please contact our support team on 08700 10 20 46. In order to help you, have the serial number of your autoclave and a detailed description of the fault report ready.
Possible Cause
Overloaded cassettes or instrument baskets.
Too many cassettes loaded into the unit.
Action
Use the heavy duty cycle for cassettes and hinged instruments.
Check filters and wash arms for debris.
Increase wash times.
The wash times for P3 and P4 can be set independently. Touch the cycle for which you want to set the wash time. Use the up and down arrows to select the desired wash time. Touch “default” if you wish to return to the factory default. Use the back arrow to return to the previous screen.
Action
Ensure that the power button is turned ON.
Ensure that the plug is properly inserted into the wall socket.
Check if a fuse or circuit is blown – call our service team on 0800 028 4749 for assistance.
Fuses should not be replaced by the operator. Call our service team on 0800 028 4749. For continued protection against the risk of fire, replace only with 15A, 250V type F fuses.
Action
Remove the wash arms, clean and reinstall them.
Check for obstructions in chamber.
Possible Cause
The cycle is not complete.
The cycle was aborted and the unit has not completed draining yet.
Action
Allow the cycle to complete.
Possible Cause
The water is not hot enough.
Action
Check that the incoming hot water temperature is in the recommended range.
Check that the hot and cold hoses are not reversed.
Action
Check for a kink in chemical tubing.
Check cap for seal.
Reinstall chemical solution by following the steps below:
Replacing the HIP™ Ultra cleaning solution
Only replace the cleaning solution with HIP™ Ultra cleaning solution, which is used in dilute form as a detergent for the cleaning of surgical instruments in the HYDRIM washer-disinfector.
It is a colourless and almost odourless liquid that is completely soluble in water. Some precautions must be taken to safely handle this chemical. For more information, see section 3.2 Handling and Installing the HIP™ Ultra cleaning solution of the Hydrim C61 Washer User Manual.
To avoid spillage, replace the solution container only when the message appears and be sure to wear gloves.
When a red X appears on the water softener/detergent icon
Press on the icon to determine if the problem is with the water softener or cleaning solution level. If the cleaning solution must be replaced, a red X will appear next to it.
To replace the cleaning solution, follow these steps:
To prime the cleaning solution dosing pump, press the water softener/detergent icon on the main screen. In the water softener/detergent screen, press the red X next to “Detergent”. The unit will prime the dosing system and a green check mark will appear in place of the red X when it is ready for use.
The system can also be primed by simply starting a cycle and selecting “Detergent Replaced” when prompted.
Action
Increase the drying time.
The drying times for P2, P3 and P4 can be set independently. Touch the cycle for which you want to set the drying time. Use the up and down arrows to select the desired drying time. Touch “default” if you wish to return to the factory default. Use the back arrow to return to the previous screen.
Change the HEPA filter by following the steps below:
To reset the drying counter:
The filter must be replaced every 750 drying cycles. The unit will continue to run if the filter is not replaced, but you will notice less than optimal drying performance.
Action
The annual maintenance should be carried out after 365 days have elapsed since the initial installation or the last service call. Call our service team on 0800 028 4749 to book an appointment.
Action
Replace HEPA Air Filter by following the steps below:
To reset the drying counter:
The filter must be replaced every 750 drying cycles. The unit will continue to run if the filter is not replaced, but you will notice less than optimal drying performance.
Action
Replace Biological Filter by following the steps below:
If filter has been replaced and the message continues to appear, reset LCS counter.
To reset the LCS counter:
Action
Check HEPA filter to ensure it is installed correctly (with arrow in correct position).
Action
If you need to copy all of the unit’s cycle data to a new USB storage device:
Action
Select the connectivity icon to go to the connection status screen.
HTM states that this must be done on a weekly basis, however always consult the manufacturer as they may request different requirements. In regards to the Scican C61 this should be done monthly.
HTM states that this must be done on a quarterly basis, however always consult the manufacturer as they may request different requirements. In regards to the Scican C61 this should be done monthly.
In respect of the Scican C61 the HIP solution should last approximately 18 cycles, depending on water quality. Your unit will inform you when these need changing.
This will be down to your individual needs. Your data logger will hold more than 7 year’s worth of data so a constant back up is achieved. It is good practice to try and download the data on a weekly basis so that the practice has a back up on a central PC for inspection purposes.
Only replace the cleaning solution with HIP™ Ultra cleaning solution, which is used in dilute form as a detergent for the cleaning of surgical instruments in the HYDRIM washer-disinfector.
It is a colourless and almost odourless liquid that is completely soluble in water. Some precautions must be taken to safely handle this chemical. For more information, see section 3.2 Handling and Installing the HIP™ Ultra cleaning solution of the Hydrim C61 Washer User Manual.
To avoid spillage, replace the solution container only when the message appears and be sure to wear gloves.
When a red X appears on the water softener/detergent icon
Press on the icon to determine if the problem is with the water softener or cleaning solution level. If the cleaning solution must be replaced, a red X will appear next to it.
To replace the cleaning solution, follow these steps:
To prime the cleaning solution dosing pump, press the water softener/detergent icon on the main screen. In the water softener/detergent screen, press the red X next to “Detergent”. The unit will prime the dosing system and a green check mark will appear in place of the red X when it is ready for use.
The system can also be primed by simply starting a cycle and selecting “Detergent Replaced” when prompted.
When a red X appears on the detergent and water softener level indicator, press on the icon to determine if the problem lies with the water softener or cleaning solution level. If the water softener must be refilled, follow these steps:
Air filter
When the message “Replace air filter” appears, the HEPA filter is in need of changing.
To reset the drying counter:
The filter must be replaced every 750 drying cycles. The unit will continue to run if the filter is not replaced, but you will notice less than optimal drying performance.
Bacteriological filter
Models fitted with a lumens cleaning system (LCS) have a biological filter that must be replaced every 500 cycles or if dirty. To change the biological filter, follow these steps:
Filter maintenance
Inspect the coarse and fine filters in the bottom of the chamber daily for debris and clean if necessary. To clean, remove the filter (turn the metal nut at the centre front of the filter to release it), rinse under a tap and reassemble. Ensure that the filter is firmly locked into position when replaced.
Wash arm maintenance
If you see that the wash arms are not turning easily, remove the arm(s) as needed – excessive removal of wash arms can reduce service life. Both the upper and lower arms are pressure mounted. To remove the upper arm, pull down and to remove the lower arm, pull up. Rinse under a tap, clear obstructions from outlet holes and reassemble.
To clean outer surfaces and the decal covering the touchscreen, use a soft cloth moistened with soap and water or a surface disinfectant. Do not use harsh cleaning chemicals. To clean the chamber, use the “Cleaning” function available in the user menu with 0.5 L of vinegar or citric acid in the chamber.
Every 25 cycles, the user will be reminded to run a chamber cleaning cycle. Press OK to continue the selected wash cycle but remember to run a chamber cleaning cycle at the next available opportunity.
If preparing the unit for transport or storage, waste water should be removed from the sump. Waste water in the unit may contain biological contaminants.
If the message “Preventative Maintenance Required. Call for Service” appears, the annual maintenance is required. To continue using the HYDRIM, press OK. Call our servicing team on 0800 028 4749 as soon as possible. The annual maintenance should be done after 365 days have elapsed since the initial installation or since the last service call was performed.
The following outlines the recommended regular maintenance to be performed by the user and technicians for the HYDRIM C61wd G4 washer-disinfector.
To ensure the unit's ongoing cleaning efficacy, SciCan recommends observing the following periodic testing protocols.
This outlines the minimum requirements to ensure that the equipment performs to specification. In some regions, local/national guidance documents may require additional testing. Please contact our Decontamination Specialists or your local authority for details.
Installation, commissioning, annual servicing must be undertaken by one of our technicians. Failure to maintain this equipment may invalidate the results of the following testing regime.
The recommended periodic testing protocols are listed in the table below:
Tests to be performed by the user:
Door lock check
This test is to ensure that the door cannot be opened while a cycle is running or a cycle will not start with the door open. To check that these features are working correctly, attempt to open the door using the handle immediately after a cycle has started.
Caution – DO NOT attempt this when the wash arms are rotating and/or the water is above ambient temperature.
The door should not open. When you attempt to start a cycle with the door open, the cycle should not start. If the unit fails this test, stop using it immediately and call our service team on 0800 028 4749.
Door seal check
Visually check the door seal for debris or wear. If the seal is worn or damaged, call our service team on 0800 028 4749.
Load carrier check
Visually check load carriers (trolleys) for wear and tear, damage and that the runners move freely over their full range.
Chamber filter check
Remove, visually check and clean as described above.
Load check
Visually inspect each load for residual soil. Washers are designed to ensure that the surfaces of instruments are free from debris that may compromise the sterilisation process. A visual check of the instruments processed in each load should be undertaken prior to sterilisation to ensure that no residual debris is present on the instrument.
If debris is still apparent, the instrument should not be considered clean and should be reprocessed through the HYDRIM again prior to sterilisation.
Some setable materials such as cements and composites, if set on the instrument will not be removed in a HYDRIM. If these materials are noticed on the instrument then abrasive removal may be required.